Example SpecimenDefinition: SpecimenDefinition Capillary Blood, 12 h fasting

Generated Narrative: SpecimenDefinition

ResourceSpecimenDefinition "single-test-serum-capillary-fasting"

typeCollected: Capillary blood specimen (SNOMED CT#122554006)

patientPreparation: 12 hour fasting (), At rest (SNOMED CT#263678003)

timeAspect: preferrably morning time

typeTested

type: Capillary blood specimen (SNOMED CT#122554006)

preference: preferred

Containers

-MaterialTypeCapDescriptionCapacityMinimumVolume[x]
*plastic (SNOMED CT#61088005)Evacuated blood collection tube with clot activator (SNOMED CT#767384004)red cap (ContainerCap#red)red cap plastic tube 2 mL2 milliliter (Details: UCUM codemL = 'mL')0.5 milliliter (Details: UCUM codemL = 'mL')

handling

temperatureQualifier: Ambient temperature ()

temperatureRange: 15-25 °C

MaxDurations

-ValueUnitSystemCode
*60minuteUnified Code for Units of Measure (UCUM)min

handling

temperatureQualifier: Refrigerated temperature ()

temperatureRange: 2-8 °C

MaxDurations

-ValueUnitSystemCode
*8hourUnified Code for Units of Measure (UCUM)h

typeTested

type: Plasma specimen (SNOMED CT#119361006)

preference: alternate

Containers

-MaterialTypeCapMinimumVolume[x]
*glass (SNOMED CT#32039001)Evacuated blood collection tube with heparin lithium and gel separator (SNOMED CT#767390000)green cap (ContainerCap#green)2 ml (Details: UCUM codemL = 'mL')

rejectionCriterion: insufficient specimen volume (RejectionCriterion#insufficient), hemolized specimen (RejectionCriterion#hemolized)

handling

temperatureQualifier: Ambient temperature ()

temperatureRange: 15-25 °C

MaxDurations

-ValueUnitSystemCode
*60minuteUnified Code for Units of Measure (UCUM)min

handling

temperatureQualifier: Refrigerated temperature ()

temperatureRange: 2-8 °C

MaxDurations

-ValueUnitSystemCode
*8hourUnified Code for Units of Measure (UCUM)h